Manual operation

38

7.3.3.

Setting the pump pressure stage

The pressure of the circulation pump is adjustable in five grades.
After setting, the VFD-Info-Display indicates the corresponding
value.

Setp.: 120.00°C
Overt.:130.00°C
Subtmp:110.00°C
>St.Pump: 4

>St.Pump:

Setting the pump pressure stage

Setting in line 4.

The value previously set appears on the DIALOG-DISPLAY
(LCD).
A flashing segment indicates that a value needs to be
entered.

Use the keypad to enter the new value

(example: 4).

Press enter

to store the selected value.

Note:
The mechanical rotation of the circulation pump produces heat that is
radiated into the bath liquid. The lowest working temperature may thus only
be reached if the pump pressure is set to grade 1.

7.4. Setting the safety temperature (with shutdown function)

Check the safety installations at least twice a year! See page 17.

TANK 150

RES 100

°C

10

30

60

130

180

220

°C

0

40

80

160

240

320

ALARM
CODE 14

ALARM!
IntAct: xx.xx°C
ExtAct: --.--°C
Temp/level alarm

This safety feature functions independent of the regulator circuit.
When the temperature of the bath liquid has reached the safety
temperature, a complete shutdown of the heater and pump is
effected.

Safety temperature in the heat exchanger >TANK<

(excess temperature protection)

 Press the key

to indicate the safety temperature value in

line 2 >TANK< on the VFD-Info-Display and using a screwdriver
simultaneously turn the setting screw to the desired value
(example: 150 °C).

Setting range:

0 °C to 320 °C

in 2 °C steps


The alarm is indicated by optical and audible signals (continuous
tone).


The error messages opposite appear on the DIALOG-DISPLAY
(LCD) and VFD-Info-Display.