JULABO LH 85 High Dynamic Temperature System User Manual

Page 55

Advertising
background image

55



Example:
 Set the external voltage or current source output for the

equivalent of 50 °C temperature setpoint.

The value adjusted and set on the external programmer is
displayed in line 4 of the DIALOG-DISPLAY (LCD) for control
purposes (Example: ExtSet: 50.0

C).

After returning the LCD display to standard display by pressing

the temperature value adjusted and set on the external

voltage or current source is displayed in line 1
(Example: EPROG 50.00

C).





20

40

60

80

100

200

300

250

150

°C

1

2 3 4 5 6 7 8 9 10

V

2 4 6 8 10

14

18 20 mA

A

B








This EPROG input enables the use of different voltage and
current values as program parameters.

"L Value" - Setting the LOW value:

1) Adjust and set the lowest desired value on the voltage or
current source resp. (Example A: 1 V).
Wait appr. 30 seconds.
2) Assign a lower temperature threshold value to this adjusted
voltage/current value by pressing the appropriate buttons on
the keypad of the instrument (Example A: 20

C ) and set by

pressing enter

.

"H Value" - Setting the HIGH value:

1) Adjust and set the highest desired value on the voltage or
current source resp. (Example A: 10 V).
Wait appr. 30 seconds.
2) Assign an upper temperature threshold value to this
adjusted voltage/current value by pressing the appropriate
buttons on the keypad of the instrument

(Example A: 200

C) and set by pressing enter

.

 Return to the standard display by pressing escape

.

Example B in the diagram serves to illustrate that the end point
values are freely selectable.

Example out of diagram A:

 Adjusting the voltage source for an output of 7.6 V!

Line 4 of the DIALOG-DISPLAY (LCD) shows the externally set
setpoint value. The instrument calculates this value from the rise
angle of the two predecided end points (in example A: 7.6 V
correspond to an external setpoint temperature of 152.0

C ).

After returning the LCD display to standard display by pressing
escape

, this value is displayed in line 1

(Example: EPROG 152.00

C).

Notice:

If this adjustment is not correctly performed at two different points, the setpoint
setting will be incorrect.