25

5.2. Tubing

Recommended tubing:
Metal tubing, triple insulated, M16x1, Temperature range -100 °C ... +350 °C

Order No.

Length

8 930 209
8 930 210
8 930 211
8 930 214

0.5

m

1.0

m

1.5

m

3.0

m


Metal tubing, insulated, M16x1, Temperature range -50 °C ... +200 °C

Order No.

Length

8 930 220
8 930 221
8 930 222
8 930 223

0.5

m

1.0

m

1.5

m

3.0

m

Pressure max.

6.0 bar at

+20 °C

4.6 bar at

+200 °C

3.8 bar at

+350 °C


Warning:

Tubing:
At high working temperatures the tubing used for temperature application and
cooling water supply represents a danger source.
A damaged tubing line may cause hot bath liquid to be pumped out within a short
time.
This may result in:
 Fire hazard
 Explosion

hazard

 Burning of skin
 Difficulties in breathing due to hot atmosphere

Safety recommendations
 Employ suitable connecting tubing.
 Make sure that the tubing is securely attached.
 Avoid sharp bends in the tubing, and maintain a sufficient distance from

surrounding walls.

 Regularly check the tubing for material defects (e.g. for cracks).