JULABO LH 85 High Dynamic Temperature System User Manual

Page 19

Advertising
background image

19

2.2

Control indicators in the header:

Heating / Cooling / Alarm / Remote control

Temperature indication Internal or External actual value

Temperature indication in °C or °F

Display for the adjusted pump pressure stage (five grades),
adjustable via the key

.

Liquid level display (five grades)
for the reservoir.

2.3

Key to swap line 3 on the VFD

ID xx

Identification no. of control module RD

L xx

Capacity in %

E xxx.xx or I xxx.xx Actual value (external or internal)

P x.xx

Pump pressure in bar

2.4

Int

S 120 .00
E 120 .00

Key to swap line 1 on the VFD

Actual value Int or Ext alternating with line 3

2.5

TAN K 150

RES 8 0

Key to indicate the safety temperature on the VFD

Line 2: TANK - Safety temperature in internal bath
Line 3: RES - Safety temperature in internal reservoir

2.6

Adjustable excess temperature protection (safety temperature)
Used for setting the safety temperature in the internal bath, called
"TANK" on the display

2.7

Adjustable safety temperature.
Used for setting the safety temperature in the internal reservoir,
called "RES" on the display

2.8

Key for automatic filling and air purge of the unit as well as the
connected external system.

3 Socket

board

3.1

REG+E-PROG

Programmer input and temperature recorder output

3.2

STAND-BY

Stand-by input (for external emergency switch-off)

3.3

ALARM

Alarm output (for external alarm signal)

3.4

EXT. Pt100

Connector for external measurement and control sensor

3.5

SERIAL

Interface RS232/RS485

3.6

RD

Connector: control cable of control module RD

Advertising