39












ALARM
CODE 13

ALARM!
IntAct: xx.xx°C
ExtAct: --.--°C
RESERVOIR

Safety temperature in the internal reservoir >RES<

This supplementary safety installation supervises and controls the
temperature of the bath liquid in the internal reservoir.

 Press the key

to indicate the safety temperature value in

line 3 >RES< on the VFD-Info-Display and using a screwdriver
simultaneously turn the setting screw to the desired value
(example: 100 °C).

Setting range:

10 °C to 220 °C

in 2 °C steps

The alarm is indicated by optical and audible signals (continuous
tone).
The error messages opposite appear on the DIALOG-DISPLAY
(LCD) and VFD-Info-Display.

 Press enter

to quit the audible signal

.

 After eliminating the malfunction, press the mains power switch

off and on again to cancel the alarm state.


Warning:

 Set the safety temperature >TANK< 15 °C above the working temperature

setpoint.

 The safety temperature value >RES<should be set at least 25 °C below the fire

point of the bath liquid used.