12

Electrical connections:
Computer interface

RS232 or RS485

or Profi

Programmer input

0 - 10 V / 0 - 20 mA

Temperatur recorder outputs

Channel 1 / 2

0 - 10 V /

Channel 3

0 - 20 mA / 4 - 20 mA

Stand-by input

External alarm device

Switching capacity max.

30 W / 40 VA

Switching voltage max.

125 V

∼/−

Switching current max.

1 A

External measurement and control sensor

Pt100, 4-lead technique


Safety installations according to IEC 61010-2-010:

Excess temperature protection

adjustable from 0 °C to 420 °C

Low liquid level protection

float switch

Liquid level indication

optical 8-graded

Rèpartition par classe selon DIN 12876-1

Classe III FL

Supplementary safety installations

High temperature warning function

optical + audible (in intervals)

Low temperature warning function

optical + audible (in intervals)

Supervision of the working sensor

plausibility control

Reciprocal sensor monitoring between
working and safety sensors

difference >100 K

Alarm indication

optical + audible (permanent)



Environmental conditions according to IEC 61 010-1:
• Use only indoor.
• Altitude up to 2000 m - normal zero.
• Ambient temperature: see Technical specifications
• Air humidity:

Max. rel. humidity 80 % for temperatures up to +31 °C,
linear decrease down to 50 % relative humidity at a temperature of +40 °C

• Max. mains fluctuations of ±10 % are permissible.
• Overvoltage category II
• Pollution degree 2

Caution:

The unit is not for use in explosive environment

Protection class according to IEC 60 529: IP21
The unit corresponds to Class I

Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes) Class A (industrial
and commercial range).