Electrical connections – JULABO HT30-M1 High Temperature Circulator User Manual

Page 58

Advertising
background image

Electrical connections

58

10. Electrical connections

Notice:

Use shielded cables only.
The shield of the connecting cable is electrically connected to the plug housing.

SERIAL

RS232/RS485 serial interface (30)
This port can be used to connect a computer with an RS232 or RS485
cable for remote control of the circulator.

Pin assignments RS232:

Pin 2

RxD

Receive Data

Pin 3

TxD

Transmit Data

Pin 5

0 V

Signal GND

Pin 7

RTS

Request to send

Pin 8

CTS

Clear to send

Pin 1; 4; 6, 9 Reserved - do not use!


Pin assignments: RS485
Pin 3

A

Pin 5

0 VD

Signal GND

Pin 8

B

Pins 1, 2, 4, 6, 7, 9

Reserved - do not use!

Accessories:

Order No.

Description

8 980 073

RS232 interface cable 9-pol./9-pol. , 2,5 m

8 900 110

USB interface adapter cable


REG+E-PROG

Programmer input / temperature recorder output (31)

Analog inputs / outputs see page 50

Pin

Signal

1 Voltage output

Channel 1

0 ... 10 V

2 Voltage output

Channel 2

0 ... 10 V

3 GND for outputs

0 V

4 Programmer input

EPROG

0 to 10 V / 0 to 20 mA

5 Current output

Channel 3

0 to 20 mA / 4 to 20 mA

6 GND for Progammer

0 V


ALARM

Alarm output (33)

(for external alarm signal)

This potential-free change-over contact is activated in case of an alarm
when pins 2 and 3 are connected.

Switching capacity

max. 30 W / 40 VA

Switching voltage

max. 125 V

∼/−

Switching current

max. 1 A

9

6

5

1

1

2

3

4

5

6

1

3

2

Advertising