JULABO HT30-M1 High Temperature Circulator User Manual

Page 52

Advertising
background image

Menu functions

52


>EPROG: Current
L Value 0.0°C
H Value 300.0°C
ExtSet 50.0°C

EPROG : 50.00°C
IntAct: 50.00°C
Power: 50 %
Control: intern

3x

Example:
• Setting a temperature of 50 °C on the external programmer!

The value adjusted and set on the external programmer is displayed in
line 4 of the DIALOG-DISPLAY (LCD) for control purposes (Example:
ExtSet: 50.0

°C).

After returning the LCD display to standard display by pressing escape

(„Setpoint“ - see page 39) this value is displayed in line 1

(Example: EPROG 50.00

°C).












EPROG : 152.00°C
IntAct: 50.00°C
Power: 100 %
Control: intern

>EPROG: Current
L Value 20.0°C
H Value 200.0°C
ExtSet 152.0°C

3x

This EPROG input enables the use of different voltage and current
values as program parameters.

„L Value“ - Setting the LOW value:

1) Adjust and set the lowest desired value on the voltage or current
source resp. (Example A: 1 V).
2) Assign a lower temperature threshold value to this adjusted
voltage/current value by pressing the appropriate buttons on the
keypad of the programmable controller (Example A: 20

°C ) and set

by pressing enter

.

„H Value“ - Setting the HIGH value:

1) Adjust and set the highest desired value on the voltage or current
source resp. (Example A: 10 V).
2) Assign an upper temperature threshold value to this adjusted
voltage/current value by pressing the appropriate buttons on the
keypad of the programmable controller

(Example A: 200

°C) and set by pressing enter

.

• Return to the standard display by pressing escape

.

Example B in the diagram serves to illustrate that the end point values
are freely selectable.

Example out of diagram A:
• Adjusting the voltage source for an output of 7.6 V!

Line 4 of the DIALOG-DISPLAY (LCD) shows the externally set setpoint
value. The programmable controller calculates this value from the rise
angle of the two predecided end points (in example A: 7.6 V correspond
to an external setpoint temperature of 152.0

°C ).

After returning the LCD display to standard display by pressing escape

, this value is displayed in line 1

(Example: EPROG 152.00

°C).

Notice:

If this adjustment is not correctly performed at two different points, the setpoint setting will be
incorrect.

20

40

60

80

100

200

300

250

150

°C

1

2 3 4 5 6 7 8 9 10

V

2 4 6 8 10

14

18 20 mA

A

B