Status messages / error messages – JULABO HT30-M1 High Temperature Circulator User Manual

Page 65

Advertising
background image

HT30-M1

65

out_par_14

xxx.xx

Minimum internal temperature in case of cascade control


Configuration
>Identif. off
Setpoint RS232
Autostart off

(**
An adjustment of these values via a serial interface only makes sense, if
the parameter for the identification is on >off<.
If the identification is adjusted (see page 38), the control parameters Xp,
Tn and Tv of a control course are determined and stored automatically.
Therefore the adjusted values are overwritten.

11.4. Status messages / error messages

The circulator sends data (including error messages) only when the computer sends a query.


Status messages

Description

00 MANUAL STOP

Circulator in "OFF" state.

01 MANUAL START

Circulator in keypad control mode.

02 REMOTE STOP

Circulator in "r OFF" state.

03 REMOTE START

Circulator in remote control mode.



Error messages

Description

-01 LOW LEVEL ALARM

Low liquid level alarm

-02 COOLING UNIT ALARM

Failure of one or both temperature sensors of the C.U.
cooling unit.

-03 EXCESS TEMPERATURE WARNING

High temperature warning "

".

-04 LOW TEMPERATURE WARNING

Low temperature warning "

".

-05 WORKING SENSOR ALARM

Working temperature sensor short-circuited or interrupted.

-06 SENSOR DIFFERENCE ALARM

Sensor difference alarm.
Working temperature and safety sensors report a
temperature difference of more than 100 K.

-07 I

2

C-BUS ERROR

Internal error when reading or writing the I

2

C bus.

-08 INVALID COMMAND

Invalid command.

-09 COMMAND NOT ALLOWED IN CURRENT

OPERATING MODE

Invalid command in current operating mode.

-10 VALUE TOO SMALL

Entered value too small.

-11 VALUE TOO LARGE

Entered value too large.