Configuration – JULABO HT30-M1 High Temperature Circulator User Manual

Page 38

Advertising
background image

Menu functions

38

8.1. Configuration


>Configuration
Control param.
Profile Start
Int.Programmer
Inputs/Outputs
Limits
Interface
Temp.Sensor

Configuration
>Identif. off
Setpoint keyb
Autostart on
Standby no
Language engl.
Degas off

By means of the configuration functions, operation of the High
Temperature Circulator can be optimized for the current application.

• Press enter

to select the configuration submenu.

• Use the up/down cursor keys

to select the desired

option. A flashing line indicates that a value needs to be entered.

• Press the P-key

to select the parameter and press enter

.

• Press escape

to return the previous menu level.


Identification

When performing an identification for the controlled system (temperature
application system), the control parameters Xp, Tn and Tv will be
automatically determined and stored.

Possible parameters:

off - no identification.

The control parameters ascertained during the last identification are
used for control purposes.

once - single identification

The High Temperature Circulator performs a single identification of
the controlled system after start.
After the identification process the parameter is automatically set to
“off“.

always - continual identification

The High Temperature Circulator performs an identification of the
controlled system whenever a new setpoint is to be reached.
NOTE: Use this setting only when the temperature application
system changes permanently.


Note:

Requirement for an identification of the controlled system:
• The High Temperature Circulator must heat to a setpoint temperature

at least 10 °C above the previous setpoint using the adjusted heating
power.