JULABO HT30-M1 High Temperature Circulator User Manual

Page 63

Advertising
background image

HT30-M1

63

in-commands: Asking for parameters or temperature values to be displayed.

Command

Parameter

Response of circulator

in_mode_03

none

Type of the programmer input:

0 = Voltage

0 V to 10 V

1 = Current

0 mA to 20 mA

in_mode_04

none

Internal/external temperature control:

0 = Temperature control in the circulator bath.

1 = Temperature control with external Pt100 sensor.

in_mode_05

none

Circulator in Stop/Start condition:

0 = Stop

1 = Start

in_mode_08

none

Adjusted control dynamics

0 = aperiodic

1 = standard

in_par_01

none

Time constant of the external bath.

in_par_02

none

Internal slope.

in_par_03

none

Time constant of the internal bath.

in_par_04

none

Band limiting (max. difference between the temperatures in the internal
bath and external system).

in_par_06

none

Xp control parameter of the internal controller.

in_par_07

none

Tn control parameter of the internal controller.

in_par_08

none

Tv control parameter of the internal controller.

in_par_09

none

Xp control parameter of the cascade controller.

in_par_10

none

Proportional portion of the cascade controller.

in_par_11

none

Tn control parameter of the cascade controller.

in_par_12

none

Tv control parameter of the cascade controller.

in_par_13

none

Adjusted maximum internal temperature in case of cascade control

in_par_14

none

Adjusted minimum internal temperature in case of cascade control