JULABO HT30-M1 High Temperature Circulator User Manual

Page 64

Advertising
background image

Remote control

64

out commands: Setting parameters or temperature values.

Command

Parameter

Response of circulator

out_mode_02

0

No identification.
Temperature control by using the stored parameters.

out_mode_02

1

Single identification of controlled system after the next start.

out_mode_02

2

Continual identification of controlled system whenever a new setpoint is
to be reached.

out_mode_04

0

Temperature control of internal bath.

out_mode_04

1

External control with Pt100 sensor.

out_mode_05

0

Stop the circulator = r OFF.

out_mode_05

1

Start the circulator.

out_mode_08

0

Control dynamics aperiodic

out_mode_08

1

Control dynamics standard

out_sp_00

xxx.x

Set working temperature "T".

out_sp_03

xxx.x

Set high temperature warning limit

.

out_sp_04

xxx.x

Set low temperature warning limit

.

out_hil_00

xxx

Set the desired maximum cooling power (0 % to 100 %).

This adjustment is required only for proportionally controlled refrigerated
circulators.

out_hil_01

xxx

Set the desired maximum heating power (10 % to 100 %).

out_par_04

xxx

Band limiting during external control.
Setting the maximum difference between the temperatures
in the internal bath and external system.

out_par_06

xxx

Xp control parameter of the internal controller. (**

out_par_07

xxx

Tn control parameter of the internal controller. (**

out_par_08

xxx

Tv control parameter of the internal controller. (**

out_par_09

xxx

Xp control parameter of the cascade controller.

(**

out_par_10

xxx

Proportional portion of the cascade controller. (**

out_par_11

xxx

Tn control parameter of the cascade controller.

(**

out_par_12

xxx

Tv control parameter of the cascade controller. (**

out_par_13

xxx.xx

Maximum internal temperature in case of cascade control