Example – JULABO HT30-M1 High Temperature Circulator User Manual
Page 57
HT30-M1
57
• Safety sensor in the cooling zone: The temperature in the cooling zone has
exceeded 170 °C. See „Important notice“ Page 25.
⇒
• Press enter
to quit the audible signal.
After eliminating the malfunction, press the mains power switch off and on again
to cancel the alarm state.
• If the unit cannot be returned to operation, contact an authorized JULABO
service station.
Mains Fuses:
• Mains circuit breakers (resettable) for the circulator (16 A) and
the CU-unit (4 A) on the rear of the housing are safety machines.
Mains Fuses:
The mains fuses (36) on the rear of the unit may easily be exchanged as
shown on the left – T 16 A.
Warning:
Before exchanging the fuses, turn off the mains power switch and disconnect the power plug
from the mains socket!
Only use fine fuses with a nominal value as specified.
Example:
Manufacturer
Supplier
Type
Order No.
Schurter
Schurter
G-fuse insert SPT
T16A 5x20mm
No. 0001.2516
60
30
15
45
4 s
16
A
M
P
4
A
M
P