Countercooling – JULABO HT30-M1 High Temperature Circulator User Manual

Page 31

Advertising
background image

HT30-M1

31

5.8. Countercooling

Caution:

Securely attach all tubing to prevent slipping.

From a working temperature of 200°C on always work with counter-cooling.





HT30

(45)

(44)

Thread M12 /

∠37 +

Barbed fittings 8 mm ID

CU

(47)

(46)

Tube connections 10 mm ID

On the HT30-circulator we distinguish two different cooling devices:
(Cooling water see page 13)

1. A cooling coil in the cooling zone for cooling water.


Without C.U.-cooling unit
The cooling wate is directly connected to the connections (44,
45).

Flow of cooling water: app. 2 l/min at

>150 °C.

or
A magnetic valve for clocked cooling water is connected to the
connection (44)
When required the control exit opens and closes the magnetic
valve.
JULABO Order no..

8 980 704

Magnetic valve


With assembled C.U.-cooling unit
the cooling water is connected to the connections (47, 46)
The flow of cooling water is controlled automatically.

HT60

Returnf (52)

HT60

Feed (53 )

Thread M16x1

2. A cooling coil in the C.U.-cooling unit for rapid cool-down of the

tempering liquid

Via the pump connections (52, 53) the tempering liquid is led
through the CU-unit.
The heat is drained via the cooling water (47).

Comments:
• With an assembled C.U.-cooling unit the menu-point CoolingMax has

to be set to a value of >0. So, the automatic control of the cooling can
become active.

• In the C.U.-unit water and oil temperature are supervised at each

outflow.
The oil flow-through is interrupted
when the oil temperature is app. 150 °C and
the water temperature is app. 80 °C.

C.U.-cooling unit:
Cooling capacity at 2 l/min flow of cooling water.

Temperature <IntAct>

[°C] 350

250

150

75

Cooling capacity

[kW]

12

8

4

1