Cooling water connection – JULABO HT30-M1 High Temperature Circulator User Manual

Page 13

Advertising
background image

HT30-M1

13

2.5. Cooling water connection

Cooling water pressure (IN / OUT )

max.

4.5 bar

Difference pressure (IN - OUT )

2.0 to 4.5 bar

Rate of flow

typical

2 l/min

Cooling water temperature

<20 °C


Notice:

Danger of corrosion of heat exchanger due to unsuitable quality of cooling water.

Due to its high content of lime, hard water is not suitable for cooling and causes scale in
the heat exchanger.

Ferrous water or water containing ferrous particles will cause formation of rust even in
heat exchangers made of stainless steel.

Chlorinated water will cause pitting corrosion in heat exchangers made of stainless steel.

Due to their corrosive characteristics, distilled water and deionized water are unsuitable
and will cause corrosion of the bath.

Due to its corrosive characteristics, sea water is not suitable.

Due to its microbiological (bacterial) components, which settle in the heat exchanger,
untreated and unpurified river water and water from cooling towers is unsuitable.

Avoid particulate matter in cooling water.

Avoid putrid water.

Recommended quality of cooling water:

pH

7.5 to 9.0

Sulfate [SO4 2- ]

< 100 ppm

Hydrocarbonate [HCO 3-]/sulfate [SO4 2-]

> 1 ppm

Hardness [Ca 2+, Mg 2+]/[HCO 3-]

> 0.5

°

dH

Alkalinity

60 ppm < [HCO 3-] < 300 ppm

Conductivity

< 500 μS/cm

Chloride (Cl -)

< 50 ppm

Phosphate (PO4 3-)

< 2 ppm

Ammonia (NH3)

< 0.5 ppm

Free chlorine

< 0.5 ppm

Trivalent iron ions (Fe 3+)

< 0.5 ppm

Manganese ions (Mn 2+)

< 0.05 ppm

Carbon dioxide (CO2)

< 10 ppm

Hydrogen sulfide (H2S)

< 50 ppm

Content of oxygen

< 0.1 ppm

Algae growth

impermissible

Suspended solids

impermissible