Cooling water connection – JULABO HT30-M1 High Temperature Circulator User Manual

Page 13

Advertising
background image

HT30-M1

13

2.5. Cooling water connection

Cooling water pressure (IN / OUT )

max.

4.5 bar

Difference pressure (IN - OUT )

2.0 to 4.5 bar

Rate of flow

typical

2 l/min

Cooling water temperature

<20 °C


Notice:

Danger of corrosion of heat exchanger due to unsuitable quality of cooling water.

•

Due to its high content of lime, hard water is not suitable for cooling and causes scale in
the heat exchanger.

•

Ferrous water or water containing ferrous particles will cause formation of rust even in
heat exchangers made of stainless steel.

•

Chlorinated water will cause pitting corrosion in heat exchangers made of stainless steel.

•

Due to their corrosive characteristics, distilled water and deionized water are unsuitable
and will cause corrosion of the bath.

•

Due to its corrosive characteristics, sea water is not suitable.

•

Due to its microbiological (bacterial) components, which settle in the heat exchanger,
untreated and unpurified river water and water from cooling towers is unsuitable.

•

Avoid particulate matter in cooling water.

•

Avoid putrid water.

Recommended quality of cooling water:

pH

7.5 to 9.0

Sulfate [SO4 2- ]

< 100 ppm

Hydrocarbonate [HCO 3-]/sulfate [SO4 2-]

> 1 ppm

Hardness [Ca 2+, Mg 2+]/[HCO 3-]

> 0.5

°

dH

Alkalinity

60 ppm < [HCO 3-] < 300 ppm

Conductivity

< 500 μS/cm

Chloride (Cl -)

< 50 ppm

Phosphate (PO4 3-)

< 2 ppm

Ammonia (NH3)

< 0.5 ppm

Free chlorine

< 0.5 ppm

Trivalent iron ions (Fe 3+)

< 0.5 ppm

Manganese ions (Mn 2+)

< 0.05 ppm

Carbon dioxide (CO2)

< 10 ppm

Hydrogen sulfide (H2S)

< 50 ppm

Content of oxygen

< 0.1 ppm

Algae growth

impermissible

Suspended solids

impermissible