Remote control

66

Error messages

Description

-12 TEMPERATURE MEASUREMENT ALARM

Error in A/D converter.

-13 WARNING : VALUE EXCEEDS

TEMPERATURE LIMITS

Value lies outside the adjusted range for the high and low
temperature warning limits. But value is stored.

-14 TEMPERATURE/LEVEL ALARM

Safety temperature alarm

-15 EXTERNAL SENSOR ALARM

External control selected, but external Pt100 sensor not
connected.

-19 TEMPERATUR MEASUREMENT ALARM

Error in the measuring system of the external temperature
sensor „EXT“

-26 STAND-BY PLUG IS MISSING

External stand-by contact is open.
Stand-by input - see pages and 6.

-31 INTERNAL COMMUNICATION ERROR

Communication between circulator HTxx and the control
electronics Mx.

-33 SAFETY SENSOR 1 ALARM

Safety sensor 1: Cable of the safety sensor 1 interrupted or
short-circuited.

-34 SAFETY SENSOR 2 ALARM

Safety sensor 2: Cable of the safety sensor 2 interrupted or
short-circuited.

-35 SAFETY SENSOR 3 ALARM

Safety sensor 3:
Cable of the safety sensor 3 interrupted or short-circuited.

-36 SENSOR COOLING ZONE ALARM

Safety sensor in the cooling zone:
Cable of the safety sensor interrupted or short-circuited.

-37 INTERNAL ALARM

Safety sensor in the cooling zone:
The temperature in the cooling zone has exceeded 170 °C!
See „Important notice Page 25