Instruction manual V 1.3 – 10/2007

4.

Setup

4.1

Overview - instrument parts / functions

Fig. 3a

1.3

1.1

1.2

1

2.1

2.2

2

1.4

1. Leica SM2500 - Microtome
1.1 Knife holder system (not part of standard delivery)
1.2 Specimen holder system (not part of standard delivery)
1.3 Carrying handle, rear (extendible - not visible in Fig. 3a)
1.4 Carrying handle, front (extendible)

2. Leica SM2500 - control unit
2.1 Control unit
2.2 Emergency stop button
3. Foot switch (not visible in Fig. 3a -

see chapter 5.4.1 for details)

4.2

Site requirements

Bench top surface for microtome and control unit sized approx. 1,000 x 950 mm

Stable laboratory bench (total weight of microtome and control unit approx. 100 kg), exactly hori-
zontal or horizontally adjustable.

At least 15 cm space on all sides of both instruments to ensure that air can circulate freely around
the instruments.

Virtually vibration-free floor.

Stable ambient temperature between +18 °C and +35 °C.

Relative air humidity of maximum 60 %, non-condensing.

Do not place near sources of vibration, avoid exposure to direct sunlight and marked temperature
fluctuations.