3 sectioning, Sectioning, Operation – Leica Biosystems SM2500 User Manual

Page 62

Advertising
background image

62

Instruction manual V 1.3 – 10/2007

5.6.3 Sectioning

Sectioning can be done in modes of operation AUTO or PROG.

Sectioning in AUTO mode of operation:

Preparatory steps:

As described for trimming (see page 59).

Select section thickness and speed according to the type of specimen being
sectioned.

Select a sectioning mode (see chapter 5.4.6). Sectioning can be done in
either CONT, SINGLE or STEP mode of operation.

Starting sectioning:

Press RUN STOP + RUN ENABLE or press the foot switch.

In SINGLE mode, the instrument will carry out one single step and then
stop in the stop or photo position selected.

In STEP mode, the instrument will continue sectioning as long as RUN
STOP + RUN ENABLE or the foot switch are being held.

To stop sectioning in CONT mode:

Press RUN STOP or RUN ENABLE or press the foot switch once again.

Sectioning in PROG mode:

Preparatory steps:

Proceed as described for trimming.

If necessary, select sectioning thickness and speed appropriate for the type
of specimen being sectioned.

Enter the desired number of sections ‘n’ you wish to be carried out (area
2.2 chapter 5.4.5).

Start sectioning:

Press RUN STOP + RUN ENABLE or press the foot switch:

The instrument will carry out the number of sections programmed and
then stop in the defined stop or photo position.