5.6.3 Sectioning

Sectioning can be done in modes of operation AUTO or PROG.

Sectioning in AUTO mode of operation:

Preparatory steps:

As described for trimming (see page 59).

•

Select section thickness and speed according to the type of specimen being
sectioned.

•

Select a sectioning mode (see chapter 5.4.6). Sectioning can be done in
either CONT, SINGLE or STEP mode of operation.

Starting sectioning:

•

Press RUN STOP + RUN ENABLE or press the foot switch.

•

In SINGLE mode, the instrument will carry out one single step and then
stop in the stop or photo position selected.

•

In STEP mode, the instrument will continue sectioning as long as RUN
STOP + RUN ENABLE or the foot switch are being held.

•

To stop sectioning in CONT mode:

Press RUN STOP or RUN ENABLE or press the foot switch once again.

Sectioning in PROG mode:

Preparatory steps:

Proceed as described for trimming.

•

If necessary, select sectioning thickness and speed appropriate for the type
of specimen being sectioned.

•

Enter the desired number of sections ‘n’ you wish to be carried out (area
2.2 chapter 5.4.5).

Start sectioning:

•

Press RUN STOP + RUN ENABLE or press the foot switch:

The instrument will carry out the number of sections programmed and
then stop in the defined stop or photo position.