Clamping cylinder-shaped specimen blocks

•

Remove screws (6) holding the fixed clamp-
ing jaw (1) (Allen key, size 5).

•

Turn the jaw and re-insert with the V-notch
facing inwards, towards the specimen.

•

Re-insert and tighten screws (6) of clamping
jaw (1).

•

Insert the movable jaw, (2) in a position that is
appropriate for the specimen size to be
clamped and with the V-notch facing inwards
(towards the specimen).

•

Tighten the movable jaw via clamping screw
(3), using an Allen key size 8 or a single-head
wrench, size 13.

Important - for square as well as for
cylindrical specimen blocks:
When the specimen block is tightly
clamped, the front edge of block (4)
must not have moved beyond the two
lateral markings (5).

Clamping specimen stages

•

The fixed clamping jaw (1) remains in standard
position (V-notch facing outwards, i.e. away
from the specimen stage).

All specimen stage sizes (plastic and metal spec-
imen stages) have a clamping block of identical
size and are clamped as follows:

•

Insert the movable jaw (2) with its front edge
(i.e. the edge facing the specimen) flush with
marking (7).

•

Insert the specimen stage (e.g. 8, 9).

•

Tighten the movable jaw via clamping screw
(3), using an Allen key size 8 or a single-head
wrench, size 13.