Cleaning, disinfection and maintenance – Leica Biosystems SM2500 User Manual

Page 69

Advertising
background image

69

Leica SM2500

Replacing the vacuum cleaner or lamp fuse

•

Insert a screwdriver into slot (1); push fuse
holder (2) slightly inwards turning at the same
time by 1/4 turn to the left.

•

The fuse holder is released and can be re-
moved.

Replacing the line fuses

•

To replace the line fuses, proceed exactly as
per chapter 4.5.1 ‘Verifying and adjusting the
voltage selector’.

•

Do not, however, change the orientation of the
voltage selector, but simply replace the defec-
tive fuses by new ones of correct specifica-
tion.

•

Remove the defective fuse (3) from the fuse
holder (2) and replace by a replacement fuse
of correct specification.

•

Reinsert the fuse holder / fuse into the instru-
ment. Use the screwdriver to push slightly in-
wards, turning at the same time by 1/4 turn to
the right.

7.

Cleaning, disinfection and maintenance

Caution!
After replacing the line fuses and rein-
serting the voltage selector housing
into the instrument, double-check
whether the correct voltage selector
setting is visible.

1

2

Fig. 76

2

3

Fig. 77