background image

4.4

Unpacking, transport, installation

Be sure to follow the unpacking instructions (attached to the out-
side of the instrument shipping crate)!

When transporting the instrument, always lock the extendible
carrying handles in place by attaching the transport safety de-
vices (white plastic blocks)!

Pull out the carrying handles (1) at the front and rear of the instrument.

Remove the upper housing parts (2) at the front and rear.

Secure the carrying handles (1) placing the white plastic blocks (3) on
the telescopic bars (4).

Two persons are needed to lift and/or carry the instrument (instru-
ment weight!).
The instrument may only be transported in an upright position!

Once the instrument has been placed onto the bench, remove the two
plastic blocks (3) from the telescopic bars (4) and put them back into
their storage place in the instrument housing.

Put the upper housing parts (2) back in place and push the carrying
handles (1) inwards as far they will go.

Prior to connecting the instrument to mains, verify whether the
voltage setting is correct for the mains electricity in your lab
(please refer to chapter 4.4.1 ‘Verifying and adjusting the voltage
selector’)!

The instrument may not be operated in hazardous locations where
risk of explosion exists!

Never remove or modify any of the safety devices installed on the
instrument and/or accessories!

4.

Setup

16

Fig. 4

1

1

2

2

3

4

Fig. 5

3

4

Fig. 6

Instruction manual V 1.3 – 10/2007