Operation – Leica Biosystems SM2500 User Manual

Page 40

Advertising
background image

40

Instruction manual V 1.3 – 10/2007

5.5.2 Clamping specimens or specimen

stages in the base plate
with vise clamp

In the base plate with vise you can ei-
ther directly clamp specimen blocks or
specimen stages holding specimen
embeddings.

Clamping square specimen blocks

The vise is equipped with two clamping jaws: a
fixed jaw (1) and a movable jaw (2) the position of
which can be changed. Both have a V-notch on
one side.

The fixed jaw (1) remains in its standard posi-
tion (V-notch facing outwards, i.e. away from
the specimen).

Insert the movable jaw (2) in a position that is
appropriate for the specimen size to be
clamped, also with the V-notch facing out-
wards (away from the specimen).

Insert the specimen block.

Tighten the movable clamping jaw via clamp-
ing screw (3), using an Allen key, size 8 or a
single-head wrench, size 13.

Important - for square as well as for
cylindrical specimen blocks:
When the specimen block is tightly
clamped, the front edge of block (4)
must not have moved beyond the two
lateral markings (5).

5.

Operation

2

1

3

5

4

5

Fig. 38