Leica SM2500

5.5.5 Specimen orientation

Caution:
When working with the eccentric ring,
NEVER attempt to orient the specimen,
as this would severely damage the
specimen orientation mechanism!

•

Releasing the base plate:

Place torque wrench (1) onto eccentric bush
(2) (arrow on the wrench must point to the
right).

•

Turn the torque wrench counterclockwise,
until the eccentric bush becomes loose.

•

Turn knurled knob (3) (marked L/R in Fig. 52)
on the left of the eccentric bush for adjustment
along the x-axis.

•

Turn knurled knob (4) (marked H/V) in Fig. 52)
on the right of the eccentric bush for adjust-
ment along the y-axis.

•

Once the specimen has been oriented, place
the torque wrench (1) back onto the eccentric
bush (2) (arrow on wrench pointing to the right)
and turn clockwise until a slight click can be
heard and felt (= tightening torque of 45 Nm).

DO NOT TURN ANY FURTHER, to pre-
vent the thread of the eccentric bush
from being stripped!