Setup – Leica Biosystems SM2500 User Manual

Page 18

Advertising
background image

18

Instruction manual V 1.3 – 10/2007

4.5.2 Installing power cord, accessory cords and connecting cable to the microtome

At the rear of the control unit, set up the following connections:

•

Remove the tape from the mains inlet for the power cord (8).

•

From the set of power cords delivered (the instrument is delivered with
several, country-specific power cords), select the one that is appropriate
for your lab; connect the power cord.

Do not yet plug the power cord into the mains outlet.

Connect the remaining cables:

•

Install the connecting cable (1) leading to the microtome and tighten the
fixing screws.

•

Connect the other end of cable (1) to the right side of the microtome and
tighten the fixing screws.

•

The connecting cable to the ultramiller (2) needs to be installed only
when working with the Leica SP2600 --> see separate instruction manual
Leica SP2600.

•

Connect photo camera (3) (optional).

•

Connect remote alarm (4) (optional).

•

Connect foot switch (5) (NOT optional).

The foot switch must be connected - otherwise the instrument will
not be operational.

•

Connect the vacuum cleaner (6) (optional with Leica SM2500 - necessary
when working with Leica SP2600).

•

Connect the lamp (7) (optional).

4.

Setup

Fig. 8

1

3

2

5

8

6

7

4

Fig.
9