Operation – Leica Biosystems SM2500 User Manual

Page 39

Advertising
background image

39

Leica SM2500

Mounting the base plate onto the sledge

•

Remove the lid (1) from the sledge.

•

Position the sledge about 3 cm away from the
front limit position:
In front, there must be enough space to easily
place torque wrench (8) on eccentric bush (2).
The microtome spindle must not be visible in
opening (3).

•

Pull eccentric bush (2) forward until the red ring
becomes visible and the red line is on top.

•

If desired, insert eccentric ring (4) (plane side
facing upwards!).

The eccentric ring blocks the orienta-
tion mechanism to ensure the section-
ing plane is not altered.

Caution!
When working with the eccentric ring,
NEVER attempt to orient the specimen,
as this would severely damage the
specimen orientation mechanism! - See
also chapt. 5 ‘Specimen orientation’.

•

Insert clamping device (6) of base plate (5) ver-
tically into opening (3).

•

Lock centering pin (7) into place (tilt slightly).

•

Push eccentric bush (2) inwards as far as it will
go (red line must be on top).

•

Turn eccentric bush (2) clockwise until the
green line is on top.

•

Place torque wrench (8) (size 32) on the eccen-
tric bush (with the arrow pointing to the right).

•

Turn torque wrench clockwise until a slight
click can be heard and felt (= tightening torque
of 45 Nm).

DO NOT TURN ANY FURTHER, to pre-
vent the thread of the eccentric bush
from being stripped!

5.

Operation

Eccentric ring - optional!

4

1

2

3

5

6

7

2

8

Fig. 34a

Fig. 35

Fig. 36

Fig. 37

Fig. 34