Mounting the specimen clamp 44 x 58 mm with
the dovetail guide located in longitudinal direc-
tion

•

Slide the clamp onto the dovetail guide so that
the three clamping screws (1) are located in
front (if necessary, relocate the three clamp-
ing screws - see below).

•

With the Allen key, size 4, tighten the two Allen
screws (2) located on the left (2).

•

Insert the specimen block and tighten the
three clamping screws (1) with the Allen key,
size 4.

Relocating the clamping screws

•

Remove the two outer clamping screws (1)
completely.

•

Use the miniature Allen key (size 1.5 - deliv-
ered together with clamp) to open Allen screw
(3).

•

Remove pressure plate (4).

•

Unscrew center clamping screw (1).

•

Insert center clamping screw (1) into the bore
(5) from the outside inwards.

•

Insert pressure plate (4) and slide onto center
clamping screw.

•

Fasten Allen screw (3) just slightly.

•

Re-insert the outer screws (1).

Specimen clamp 44 x 58 mm mounted correctly onto the
dovetail guide installed in longitudinal direction

Specimen clamp 44 x 58 mm - with clamping screws
located correctly for installation on dovetail guide
located in
longitudinal direction