43

Leica SM2500

Mounting the specimen clamp 44 x 58 mm with
the dovetail guide located in longitudinal direc-
tion

Slide the clamp onto the dovetail guide so that
the three clamping screws (1) are located in
front (if necessary, relocate the three clamp-
ing screws - see below).

With the Allen key, size 4, tighten the two Allen
screws (2) located on the left (2).

Insert the specimen block and tighten the
three clamping screws (1) with the Allen key,
size 4.

Relocating the clamping screws

Remove the two outer clamping screws (1)
completely.

Use the miniature Allen key (size 1.5 - deliv-
ered together with clamp) to open Allen screw
(3).

Remove pressure plate (4).

Unscrew center clamping screw (1).

Insert center clamping screw (1) into the bore
(5) from the outside inwards.

Insert pressure plate (4) and slide onto center
clamping screw.

Fasten Allen screw (3) just slightly.

Re-insert the outer screws (1).

Specimen clamp 44 x 58 mm mounted correctly onto the
dovetail guide installed in longitudinal direction

Specimen clamp 44 x 58 mm - with clamping screws
located correctly for installation on dovetail guide
located in
longitudinal direction

5.

Operation

Fig. 45

1

12

12

12

12

12

12

1234567

KNIFE

2

4
3

1

3

4

5

Fig. 46