3.3 SDS-PAGE Buffers

Table 3.2. Recommended Criterion precast gels and buffers for SDS-PAGE.

Gel Type

Sample Buffer

Running Buffer

Criterion TGX

Laemmli (catalog #161-0737)

Tris/glycine/SDS

Criterion TGX Stain-Free

Optional: 2-mercaptoethanol

(catalog #161-0732)

Criterion Tris-HCl

(catalog #161-0710) or DTT

Criterion Stain Free

(catalog #161-0611)

Criterion XT Bis-Tris

XT sample buffer (catalog #161-0791)

XT MES (catalog #161-0789)

Optional: XT reducing agent

XT MOPS (catalog #161-0788)

(catalog #161-0792)

Criterion

XT

Tris-acetate

XT

Tricine

(catalog

#161-0790)

3.3.1 Running Buffers

See Appendix B for buffer formulations. Do not adjust pH.

3.4 Sample Preparation

1. Determine the appropriate concentration of sample to load (depends on the load volume and the

detection method used; see Chapter 12 for approximate stain sensitivities).

2. Dilute the sample with at least an equivalent volume of sample buffer with added reducing agent.

For nonreducing conditions, omit the reducing agent.

TGX and Tris-HCl Gels

XT Gels (Bis-Tris and Tris-Acetate)

4.75 μl Laemmli sample buffer

5 μl XT sample buffer

0.25 μl

β-mercaptoethanol

1 μl XT reducing agent

5 μl sample

1–14 μl sample

10 μl total volume

Make up to 20 μl total volume with diH

2

O

3. Heat the diluted sample at 90–95°C for 5 min, or at 70°C for 10 min.

Tris/glycine/SDS (1x)

25 mM Tris, 192 mM glycine, 0.1% SDS

(pH 8.3)

Dilute 100 ml 10x stock (catalog #161-0732) with 900 ml diH

2

O

XT MES (pH 6.4)

Dilute 50 ml 20x stock (catalog #161-0789) with 950 ml diH

2

O

XT MOPS (pH 6.9)

Dilute 50 ml 20x stock (catalog #161-0788) with 950 ml diH

2

O

XT Tricine (pH 8.2)

Dilute 50 ml 20x stock (catalog #161-0790) with 950 ml diH

2

O

3.3.2 Sample Buffers
Laemmli

62.5 mM Tris-HCl, pH 6.8, 2% SDS, 25% glycerol, 0.01% (catalog #161-0737)

bromophenol blue, 5%

β-mercaptoethanol or 100 mM DTT (added fresh)

XT

Use XT sample buffer (catalog #161-0791) and XT reducing agent

(catalog

#161-0792)