Troubleshooting

14

Problem

Cause

Solution

General Troubleshooting Tips

Current is zero or less than
expected, and samples do not
migrate into gel

Tape at bottom of cassette not removed Remove tape

Insufficient buffer in integral buffer
chamber

Fill buffer chamber with running buffer

Insufficient buffer in lower buffer
chamber

Fill both halves of lower buffer tank with
400 ml running buffer when running two
gels

Electrical disconnection

Check electrodes and connections

Gels run faster than expected

Running buffer too concentrated or
incorrect

Check buffer composition

Gel temperature too high

Do not exceed recommended running
conditions

Bands “smile” across gel: band
pattern curves upward at both
sides of the gel

Excessive heating of gel

Check buffer composition

Do not exceed recommended running
conditions

Insufficient buffer

Fill both halves of lower buffer tank with
400 ml running buffer when running two
gels

Bands “smile” or “frown” within gel
lanes

Protein load too high

Load less protein

Sample or buffer preparation issues

Minimize salts, detergents, and solvents
in sample preparation and sample loading
buffers

Incorrect running conditions

Set correct voltage

Bands are skewed or distorted;
lateral band spreading

Too much salt in samples

Remove salt from samples (dialysis,
precipitation, or other method)

Insufficient or wrong sample buffer

Check buffer composition and dilution
instructions

Sample precipitation

Selectively remove predominant proteins in
sample

Dilute sample in sample buffer

Insoluble materials (for example, cell
membranes) in samples

Centrifuge samples to remove particulates
prior to sample loading

38

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

Table 14 .1 . Troubleshooting electrophoresis and detection with Criterion

gels . For more troubleshooting tips, refer to the

Criterion cell, Criterion blotter, and Trans-Blot

®

SD cell instruction manuals, or contact Technical Support.