Problem

Cause

Solution

Artifactual bands at
60–70 kD

Skin keratin contamination

Clean all dishware, and wear gloves while
handling and loading gels

Filter all solutions (0.2–0.45 µm filter)

Poor resolution or fuzzy bands

Sample volume too high

If possible, load a more concentrated
sample in a lower sample buffer volume

Not enough running buffer in lower
chamber

Add running buffer to fill line of lower buffer
chamber

Diffuse sample loading zone

Load sample with a syringe or gel loading
pipet tip

Sample diffusion during staining

Fix gel with 40% methanol, 10% acetic acid
for 80 min prior to staining

Incompatible sample components

Minimize salts, detergents, and solvents
in sample preparation and sample loading
buffers

Expired gel

Use gels before expiration date printed on
the cassette

Criterion

TGX Stain-Free

and Criterion Stain Free

Gels

Low sensitivity for proteins

Low tryptophan content in proteins

After activation and imaging, stain gel with
Bio-Safe

Coomassie or similar to detect

missing bands

Uneven sensitivity or fuzzy bands

Gel was soaked in water or buffer prior
to activation and imaging

If possible, activate and image gel using
the Gel Doc EZ imager immediately after
electrophoresis

Bands are too light or missing from
blot (membrane)

Proteins transferred through membrane

Use membrane with smaller pore size

Decrease transfer time

Decrease voltage

Standards not visible

Incorrect standards were used

Use unstained standards; some prestained
standards are not detected by the Gel Doc
EZ imager. To monitor electrophoresis, use
a 1:1 mixture of unstained and prestained
standards

Dye front at bottom of gels
interfering with detection of proteins

Sample constituents present in gel
interfering with imaging

Dilute sample in gel running buffer prior to
loading

Activate and image gel, rinse in fixation
solution for 30 min, and repeat imaging

High background or low sensitivity
on a blot of a Criterion TGX Stain-
Free or Criterion Stain Free gel
(imaged using the Gel Doc EZ
imager)

Low tryptophan content in proteins

After activation and imaging, stain blot with
a total protein blot stain to detect missing
bands

Membrane not low-fluorescence PVDF

Always use low-fluorescence PVDF
membranes to blot Criterion TGX Stain-Free
or Criterion Stain Free gels

PVDF membrane is dry

Wet PVDF membrane briefly in methanol
and wash in water for 1 min before imaging