8.4 Sample Preparation
1. Determine the appropriate protein concentration of your sample based on the detection method
and load volume used. (See Chapter 12 for approximate stain sensitivities.)
2. Dilute 1 part sample with 1 part sample buffer. Do not heat the samples.
8.5 Running Conditions
Power conditions
125 V constant
Starting current
90–120 mA/gel
Final current
35–55 mA/gel
Run time
90 min
Criterion Precast Gels