4 sample preparation, 5 running conditions – Bio-Rad Criterion™ TBE-Urea Precast Gels User Manual

Page 26

Advertising
background image

20

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

6.4 Sample Preparation

1. Determine the appropriate concentration of sample to load (depends on the load volume and the

detection method used; see Chapter 12 for approximate stain sensitivities).

2. Dilute the sample with at least an equivalent volume of sample buffer (catalog #161-0739) and

reducing agent (

β-mercaptoethanol, for example). Heat the diluted sample at 90–95°C for 5 min,

or at 70°C for 10 min.

For example, combine:

5 μl sample

4.75 μl Tricine sample buffer (catalog #161-0739)

0.25

μl

β-mercaptoethanol (catalog #161-0710)

10

μl

total

volume

6.5 Running Conditions

Power conditions

125 V constant

Starting current

140–150 mA/gel

Final current

60–70 mA/gel

Run time

120 min