Appendix a: quick start guide, Quick start guide – Bio-Rad Criterion™ TBE-Urea Precast Gels User Manual

Page 46

Advertising
background image

Quick Start Guide

A

Running buffer (1x)

Dilute 100 ml 10x stock (catalog #161-0732) with 900 ml diH

2

O.

Sample buffer

Use Laemmli sample buffer (catalog #161-0737).

Prepare Buffers

Prepare and Load Samples

Prepare Gels and

Assemble Electrophoresis Cell

Perform Electrophoresis

Remove the comb and tape from the gels and assemble the electrophoresis cell.

Fill the inner and outer buffer chambers with running buffer.

Fill the lower buffer chamber with 400 ml running buffer (to the fill mark).

40

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

The following instructions are for electrophoresis of Criterion

precast gels using the Criterion system.

Component

Reducing

Nonreducing

Sample

5 μl

5 μl

Laemmli sample buffer
(catalog #161-0737)

4.75 μl

5 μl

β-Mercaptoethanol

0.25 μl

Total volume

10 μl

10 μl

Heat samples at 90–100°C for 5 min or at 70°C for 10 min.

Load the appropriate amount of sample on the gel.

Connect the electrophoresis cell to the power supply and perform electrophoresis according to
the conditions in Table A.1.

Advertising