3.5 Running Conditions

Run conditions and times are approximate. Conditions may vary depending on water and buffer
conductivity, which vary from one lab setting to the next. Multiply current by the number of gels run.

Table 3.3. Running conditions for SDS-PAGE with Criterion gels in the Criterion cell. Do not run different gel
formulations at the same time.

TGX Tris-HCl

Bis-Tris

Tris-Acetate

Running buffer

Tris/glycine/SDS

Tris/glycine/SDS

XT MOPS

XT MES

XT Tricine

Standard Conditions
Power conditions

200 V constant

200 V constant

200 V constant

200 V constant 150 V constant

Expected current (per gel)

Initial

55–80 mA

90–120 mA

165–175 mA

185–200 mA

170–180 mA

Final

33–43 mA

35–55 mA

60–70 mA

90–110 mA

85–95 mA

Run time

42–45 min

50–55 min

60 min

45 min

65 min

High Voltage (Rapid) Conditions
Power conditions

300 V constant

Expected current (per gel)

Initial

89–135 mA

Final

66–99 mA

Run time

20–26 min

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com