Electrophoresis troubleshooting – Bio-Rad Gel Doc™ EZ System User Manual

Page 36

Advertising
background image

Criterion Stain Free

TM

System | Criterion Stain Free Gels

22

Electrophoresis Troubleshooting

Improper storage of Criterion gels can produce numerous artifacts. Criterion
gels should be stored flat at 4ºC.

Avoid freezing or prolonged storage above 4ºC. If you suspect your gels have
been stored improperly, DO NOT USE THEM.

Problem

Possible Cause

Solution

Samples do not migrate
into gel

Tape at the bottom of the
cassette not removed

Remove tape

Insufficient buffer in integral
buffer chamber

Fill integral buffer chamber with 60 ml
running buffer

Insufficient lower electrode
buffer

Fill both halves of the lower buffer tank with
400 ml running buffer when running two
gels

Electrical disconnection

Check electrodes and connections

Bands “smile” across gel,
band pattern curves
upward at both sides of the
gel

Excess heating of gel

Check buffer composition

Completely fill both halves of the lower
buffer tank with 400 ml running buffer when
running two gels

Do not exceed recommended running
conditions

Skewed or distorted bands,
lateral band spreading

Excess salt in samples

Remove salts from sample by dialysis or
desalting column prior to sample
preparation

Insufficient sample buffer or
wrong formulation

Check buffer composition and dilution
instructions

Vertical streaking

Overloaded samples

Dilute sample

Selectively remove predominant protein in
the sample

Sample precipitation

Centrifuge samples to remove particulates
prior to sample loading

Gels run too fast, provide
poor resolution, and gel
temperature is too high

Running buffer is too
concentrated

Check buffer composition

Artifact bands at
60-70 kD

Possible skin keratin
contamination

Clean all dishware and wear gloves while
handling and loading gel

Filter all solutions through nitrocellulose

Use 10% iodoacetamide to eliminate keratin
bands