Appendix A: Troubleshooting

Follow the recommendations below to troubleshoot problems with the Imager.

Problem

Possible Cause

Solution

Green LED on front panel is
off

AC power cord is not
connected

Connect AC power cord to the imager and an
appropriate wall socket

Power is off

Turn the power switch on

Front panel LEDs remain
flashing

Firmware is not loading

Call Bio-Rad Technical Support for help

Image is not visible on the
monitor

No sample on sample tray

Place a sample on the tray and run protocol
again

Image is not bright enough Underdeveloped gel

If initially stopped after 2 minutes, allow the gel
to develop for 2-3 more minutes

Red LED flashing on front
panel

Sample tray not inserted all
the way

Push sample tray into the imager until you feel
the magnet pull the sample tray

Door interlock broken

Please close the door

Bulb fault

Replace all the lamps according to the
instructions in Chapter 2

Communication interrupted Make sure Imager power is on the USB cable is

connected to the PC. Restart the imager

Imaging button is pressed
on front panel while
software is not yet open

Red light will stop flashing when the software is
opened. Press the imaging button again to
initiate imaging