2.5.2 Setup Editor Screen

The Setup screen selects devices to be included in a method. BioLogic DuoFlow software versions 3.0 and
later includes the QuadTec Detector. Use the Detectors button in the Setup screen to display the Edit
Detector window. Select the QuadTec button and enter the desired wavelengths for your method. The Time
Constant value smooths the signal with a time filter. For most routine applications, a Time Constant value of
1.0 seconds is recommended.

Figure 2-8. Edit Detector Window

2.5.3 Run Screen and Chromatogram Settings

The Run screen, shown in Figure 2-10, reflects the real-time sample run. The traces shown on the Run
screen are chosen in the Chromatogram Settings window. From the system toolbar, select the Settings
button to display the Run Screen Chromatogram Settings, shown in Figure 2-9. Select the traces displayed,
depending on the instruments being used, select up to eight traces from the following: Conductivity, standard
optics module UV, GP-pressure, four QuadTec traces, two SIM traces, or two pH trace, by the Trace Device
drop-down menu in this window. Select the Visible checkbox to indicate whether or not the trace is to be
visible on the screen.

Figure 2-9. Run Screen Chromatogram Settings

USE WITH THE BIOLOGIC DUO-FLOW SYSTEM

SYSTEM INSTALLATION AND SETUP

2-8

Ok

Cancel

Detector

280

1

1.0

Time Constant

Edit Detector

QuadTec

UV Detector

SIM/pH

Set wavelength(s)
Range 190 to 740 nm

SIM Signal

260

2

214

3

405

4

UV Detector

Ok

Cancel

Default

AU

-0.000

Settings for the Run Mode

Y-axis settings

Trace Device

Visible

to-Scroll

Scrolling Time (min):

3

Units

Baseline

2.000

Axis Max

QuadTec-1 (214 nm)

QuadTec-2 (260 nm)
QuadTec-3 (280 nm)
QuadTec-4 (405 nm)
SIM1/pH
Conductivity
GP-pressure
%Buffer B

x

x
x
x
x
x

x

Chromatogram Settings

QuadTec-1 (214 nm)

x