3.1 ELECTRICAL CONNECTIONS TO THE BIOLOGIC HR SYSTEM
Set up the BioLogic HR system with the QuadTec detector on the bottom tray. Do not turn power on.
1.
Connect one SIM-HR to the HR Workstation with System Cable 17, and ensure that the SIM’s
device address number (Device Number) is set to 1.
2.
"Daisy-chain" a second SIM-HR to the first SIM-HR using a second System Cable 17. The device
address number (Device Number) of this second SIM-HR must be set to 0.
3.
Connect the QuadTec to the SIM-HR modules using two System Cable 24, an analog cable that has
red and black bare wires at one end and a gold colored banana plug at the other. Plug the banana
plugs into the 1V Integrator sockets on the rear of the QuadTec marked 1 and 2. The cable in socket
1 should go to the first SIM-HR (Device Number 1) and the cable in socket 2 goes to the second
SIM-HR (Device Number 0). In both cases the cable red wire goes to the analog + connector and
the black wire to the analog - connector of the SIM-HR.
4.
If a Model 2128 Fraction Collector is to be used, connect it to SIM 2 using System Cable 17.
Figure 3-2. Connections to the BioLogic HR System
USE WITH THE BIOLOGIC HR SYSTEM
SYSTEM INSTALLATION AND SETUP
3-2
WORKSTATION
MODEL 2128
FRACTION
COLLECTOR
SIM 2 (ADDRESS 0)
SIM 1 (ADDRESS 1)
MODEL 1327
CHART RECORDER
SYSTEM CABLE 17
SYSTEM
CABLE 17
SYSTEM CABLE 24
SYSTEM CABLE 24
SYSTEM
CABLE 2
AV7-3
AUTOMATED
INJECT VALVE
CONDUCTIVITY
MONITOR
SYSTEM
CABLE 17
SYSTEM
CABLE 4
QUADTEC
DETECTOR
COLOR MONITOR
SYSTEM CABLE 17
CONTROLLER
KEYBOARD
MOUSE