3.1 ELECTRICAL CONNECTIONS TO THE BIOLOGIC HR SYSTEM

Set up the BioLogic HR system with the QuadTec detector on the bottom tray. Do not turn power on.

1.

Connect one SIM-HR to the HR Workstation with System Cable 17, and ensure that the SIM’s
device address number (Device Number) is set to 1.

2.

"Daisy-chain" a second SIM-HR to the first SIM-HR using a second System Cable 17. The device
address number (Device Number) of this second SIM-HR must be set to 0.

3.

Connect the QuadTec to the SIM-HR modules using two System Cable 24, an analog cable that has
red and black bare wires at one end and a gold colored banana plug at the other. Plug the banana
plugs into the 1V Integrator sockets on the rear of the QuadTec marked 1 and 2. The cable in socket
1 should go to the first SIM-HR (Device Number 1) and the cable in socket 2 goes to the second
SIM-HR (Device Number 0). In both cases the cable red wire goes to the analog + connector and
the black wire to the analog - connector of the SIM-HR.

4.

If a Model 2128 Fraction Collector is to be used, connect it to SIM 2 using System Cable 17.

Figure 3-2. Connections to the BioLogic HR System

USE WITH THE BIOLOGIC HR SYSTEM

SYSTEM INSTALLATION AND SETUP

3-2

WORKSTATION

MODEL 2128

FRACTION

COLLECTOR

SIM 2 (ADDRESS 0)

SIM 1 (ADDRESS 1)

MODEL 1327

CHART RECORDER

SYSTEM CABLE 17

SYSTEM

CABLE 17

SYSTEM CABLE 24

SYSTEM CABLE 24

SYSTEM
CABLE 2

AV7-3
AUTOMATED
INJECT VALVE

CONDUCTIVITY

MONITOR

SYSTEM
CABLE 17

SYSTEM
CABLE 4

QUADTEC

DETECTOR

COLOR MONITOR

SYSTEM CABLE 17

CONTROLLER

KEYBOARD

MOUSE