Bio-Rad BioLogic QuadTec™ Detector and Components User Manual

Page 7

Advertising
background image

1.1 USING THIS MANUAL

The instructions provided in this manual are for use of the QuadTec Detector with the following systems:

The BioLogic DuoFlow chromatography system

The BioLogic HR chromatography system

Stand-alone use

1.2 OPTICAL PATH OF THE QUADTEC UV/VIS MULTI-WAVELENGTH DETECTOR

The light emitted from the lamp [1] is focussed by a lens [2] and a slit [4]. If a deuterium lamp is used and
wavelengths >380 nm are chosen, an automatic edge filter [3] is employed. The light passes to a grating
monochromator [6] for selection of the desired wavelength. The beam is then split by a semi-transparent
mirror [7]. Part of the beam passes to the photo diode reference signal [8] and optical absorption is
measured. The other part of the beam passes through the flowcell and sample and the light intensity is
measured at the sample photo diode [10]. The difference or the ratio of the two signals is calculated to
provide information about sample components.

Figure 1-2. Theory of Operations

INTRODUCTION

SYSTEM OVERVIEW

1-2

1. LAMP (DEUTERIUM OR HALOGEN)

2. LENS SYSTEM (QUARTZ OPTICS)

4. SLIT

5. CONCAVE MIRROR

6. MOVABLE GRATING

MONOCHROMATOR

10. SAMPLE PHOTO DIODE

3. EDGE FILTER

7. SEMI-TRANSPARENT MIRROR

9. FLOW CELL

8. REFERENCE PHOTO DIODE