Appendix a specifications – Bio-Rad BioLogic QuadTec™ Detector and Components User Manual

Page 52

Advertising
background image

APPENDIX A
SPECIFICATIONS

Lamp type

Deuterium (standard)
Halogen (option)
User-replaceable

Wavelength range

190–740 nm with deuterium lamp, automatic edge filter used at 380 nm
370–740 nm with halogen lamp

No. of simultaneous

4 with Biologic DuoFlow system

wavelengths

2 with Biologic HR system
2 in stand-alone mode

Bandwidth

8 nm

Wavelength accuracy

1 nm

Measurement range

0–4 AU

Sensitivity

2 x 10-5 AU at 240 nm and time constant 1.0 s

Noise

1: 1 x 10-5 AU at 240 nm, 1.0 s
2: 2 x 10-5 AU at 240/300 nm, 1.0 s

Baseline drift

1: 10 x 10-5 AU/hr at 240 nm, 1.0 s
2: 15 x 10-5 AU/hr at 240/300 nm, 1.0 s

Linearity

2% deviation up to 2.0 AU at 254 nm with acetone

Time constants

0.1/0.2/0.5/1.0/2.0/5.0/10.0 s

Auto zero range

Full scale

Weight

5.5 kg (15 lbs)

Dimensions

160 x 185 x 340 mm (W x H x D), (6.3 x 7.3 x 13.4 in)

Qualification (GLP)

Records total operating hours, lamp changes, lamp ignitions

support

Flow cell

PEEK, 3 mm pathlength, 2 µL volume

SPECIFICATIONS

APPENDIX A

A-1