Use the Right arrow key to access the output field and select E1 or E2 as
appropriate.

Use the right arrow key to access the signal level. Set the minimum signal level that
must be exceeded before fraction collection starts.

Both functions, level and event, are evaluated in a timetable to control fraction
collection. This means, that the E1 or E2 output is activated only when the level is
exceeded and during a time window in the timetable of a method. In case of missing
time programs, the appropriate event has to be set to “1” in the operation screen.

If fraction collection is controlled by the signal level only, the event output becomes
active every time the signal exceeds the preset signal level.

4.2.2 LINK Menu

The Link menu is used when running links, which are a pre-defined run of program combinations. In this
menu the momentary status of the program links can be monitored.

In the first line of the menu the momentary status can be observed, and the number of runs (R-xx) to be
performed counts downwards. In the second line the next step of the link to be executed is shown. Arrows at
the right side of the display indicate that more program lines above and below, respectively, do exist.

Figure 4-6. Overview of the Link Menu

USE AS A STAND-ALONE SYSTEM

SYSTEM INSTALLATION AND SETUP

4-5

Link

View

No LINK Running

Link

View

LINK is Running