0 specifications, Electrical specifications – ETS-Lindgren HI-1710A Microwave Oven Survey Meter User Manual

Page 17

Advertising
background image

www.ets-lindgren.com

Specifications

17

3.0 Specifications

Electrical Specifications

Ranges:

Four ranges:

0

–1, 0–2, 0–5, and 0–10 mW/cm2

Response Characteristics:

0 to 90% of final value for step

input:

F0, Raw (Version 3.XX only)

unfiltered

F1, Slow

—less than three

seconds

F2, Fast

—less than one second

Maximum Power Density:

2.0 Watts/cm

2

(2000 mW/cm

2

)

Power Input:

100

–240 VAC

0.8

–0.4 A max

50/60 Hz

Fuses:

250 Volt, 1.0 Amp, Type T

(5 mm x 20 mm)

Control Inputs:

Zero

Dry switch closure to zero

the HI-1710A

RS-232

Standard 9-pin, 2-way

communications input

Control Outputs:

Alarm Output

Form C contact;

24 V 1 Amp

Analog Output

4 VDC max;

1 mA max

RS-232

Standard 9-pin, 2-way

communications output

Advertising