Calibration, Power cable – ETS-Lindgren HI-1710A Microwave Oven Survey Meter User Manual
Page 19

www.ets-lindgren.com
Specifications
19
Calibration
Calibration:
Calibrated at 2450 MHz for use in
the ISM band
Calibration is performed at a standard temperature of 75°F in an
anechoic chamber, utilizing a slot radiator.
A CW field is generated using a crystal-controlled solid state source.
The power monitoring system is the HI-2795 Calibration Comparison
System.
The calibration field level is established using a reference survey meter
with CDRH calibration correction factors.
All ranges are calibrated at mid-scale.
Power Cable
The HI-1710A Microwave Measurement System is shipped with a three-wire
power cable. When the cable is connected to an appropriate AC power source, it
connects the chassis to earth ground. The type of power cable shipped with each
HI-1710A depends on the country of destination.