Calibration, Power cable – ETS-Lindgren HI-1710A Microwave Oven Survey Meter User Manual

Page 19

Advertising
background image

www.ets-lindgren.com

Specifications

19

Calibration

Calibration:

Calibrated at 2450 MHz for use in

the ISM band

Calibration is performed at a standard temperature of 75°F in an

anechoic chamber, utilizing a slot radiator.

A CW field is generated using a crystal-controlled solid state source.

The power monitoring system is the HI-2795 Calibration Comparison

System.

The calibration field level is established using a reference survey meter

with CDRH calibration correction factors.

All ranges are calibrated at mid-scale.

Power Cable

The HI-1710A Microwave Measurement System is shipped with a three-wire

power cable. When the cable is connected to an appropriate AC power source, it

connects the chassis to earth ground. The type of power cable shipped with each

HI-1710A depends on the country of destination.