38

Menu Options, Settings, Command Set

www.ets-lindgren.com

Mode 2

—Prints summary ticket. This is the default mode. The

HI-1710A sends a pre-formatted summary report through the RS-232

serial port each time the probe is zeroed. This is most useful when

using Peak Hold mode. With this configuration, a serial input printer set

up to print adhesive-backed labels can be connected to the RS-232

serial port. After scanning an oven, the meter is zeroed. On zeroing, a

ticket is printed with the maximum leakage observed as well as several

system test parameters.

The output is formatted for a standard 3 1/2-in x 15/16-in label

(for example, Avery® label 4013). Line feed characters are

transmitted following the formatted data to advance one additional

label space with each formatted output. By printing the report on

every other label, the labels can be removed from the backing

more conveniently.

Mode 3

—Does not print summary ticket when zeroed. Changes the

output format of the status messages S0

–S8. The transmitted

summary messages are shortened for easier adaptation to a computer

interface program.

The zero command does not transmit a formatted report as in

mode 2. In mode 3 the transmitted status messages end with only

a carriage return terminator (CR). In modes 0, 1, and 2, the

messages end with a carriage return and a line feed (CR/LF).