Menu Options, Settings, Command Set

31

AUDIO

Turns on the audible alarm indicator, which will sound to indicate that

the measured leakage has exceeded a preset level (alarm set point).

Although the alarm set point may be specified, the alarm will

not sound unless the audio is turned on. For set point

information, see page 30.

Menu Option

Setting

Serial Port

Command

OFF

Q0

ON

Q1

Q?

AUTO-RANGE

When set to auto-range, the best range for each measuring condition

will automatically be selected.

Menu Option

Setting

Serial Port

Command

Function

A0

OFF

A1

ON

A?

Retrieve status

The auto-range setting only affects the bar graph display

and the analog output.

When in Peak Hold with the range set to AUTO, the range

will change as the field strength increases; it will remain in

the range that shows the highest peak value until the ZERO

button is pressed. This is so the peak value can be

displayed on the bar graph. When the ZERO button is

pressed with no field present, the range will revert back to

the lowest (1 mW/cm2).