Menu Options, Settings, Command Set

37

PRINT

MODE

There are four modes that specify the content and length of the

information provided during testing.

See page 42 for the status message formats for modes

0, 1, and 2, and page 43 for mode 3.

Menu Option

Setting

Serial Port

Command

0

M0

1

M1

2

M2

3

M3

M?

Mode 0

—Print mode off. Select this mode when no automatic output

from the HI-1710A is needed; for example, when all output is

specifically requested by the controller. This mode might be chosen in

a fully-automatic test system where the external device requests the

measurement, data gathering, and the zeroing of the meter.

Status messages are transmitted in the same form as in mode 2,

but no message is transmitted by the HI-1710A when the zero

command is received.

Mode 1

—Prints continuous reading. In this mode the HI-1710A

continually sends the RF reading through the RS-232 serial port. Data

is transmitted at a high rate and can only be used at baud rates of

4800 and above; data is sent at a rate of 45 Hz, the same rate at which

the probe is sampled. The controlling device must be able to receive a

serially-transmitted 4-digit number 45 times a second.

This mode might be used in the laboratory to investigate the

effects of different filter constants and scan rates. The output from

the HI-1710A is a three- or four-digit ASCII string in the format

X.XX or X.XXX.