Parameter screen – ETS-Lindgren HI-1710A Microwave Oven Survey Meter User Manual

Page 35

Advertising
background image

www.ets-lindgren.com

Menu Options, Settings, Command Set

35

PARAMETER

SCREEN

Displays system board voltages and probe voltages and offsets.

Change the setting to ON, and exit the menu to display the

parameter screen. Press the ZERO button to escape the

parameter screen.

Menu Option

Setting

Serial Port

Command

OFF

ON

Sample Parameter Screen

RF

—Probe RF level currently detected.

BIAS

—Probe bias reading; value should be between 0.30 and

0.40 volts.

REF

—Power supply reference voltage; value should be between 1.22

and 1.26 volts.

+5SUP

—Power supply +5 voltage; value should be between 4.75 and

5.25 volts.

+8SUP

—Power supply +8 voltage; value should be between 7.60 and

8.40 volts.

-8SUP

—Power supply -8 voltage; value should be between -7.92 and

-8.08 volts.

OFST

—Probe offset reading; value should be between 0.02 and

0.40 mW (0.005 and 0.100 volts).