Parameter screen – ETS-Lindgren HI-1710A Microwave Oven Survey Meter User Manual
Page 35

www.ets-lindgren.com
Menu Options, Settings, Command Set
35
PARAMETER
SCREEN
Displays system board voltages and probe voltages and offsets.
Change the setting to ON, and exit the menu to display the
parameter screen. Press the ZERO button to escape the
parameter screen.
Menu Option
Setting
Serial Port
Command
OFF
—
ON
—
Sample Parameter Screen
RF
—Probe RF level currently detected.
BIAS
—Probe bias reading; value should be between 0.30 and
0.40 volts.
REF
—Power supply reference voltage; value should be between 1.22
and 1.26 volts.
+5SUP
—Power supply +5 voltage; value should be between 4.75 and
5.25 volts.
+8SUP
—Power supply +8 voltage; value should be between 7.60 and
8.40 volts.
-8SUP
—Power supply -8 voltage; value should be between -7.92 and
-8.08 volts.
OFST
—Probe offset reading; value should be between 0.02 and
0.40 mW (0.005 and 0.100 volts).