Back panel controls and connectors, Zero, Analog out – ETS-Lindgren HI-1710A Microwave Oven Survey Meter User Manual

Page 23

Advertising
background image

www.ets-lindgren.com

Operation

23

Back Panel Controls and Connectors

HI-1710A Back Panel

ZERO

Can be used to attach an external device to the ZERO connector. See

About Zeroing on page 21 for more information on the zero function.

ANALOG

OUT

This analog output BNC connector is a DC voltage varying from 0 to 4 volts

proportional to the full scale range setting. For example, with the instrument on

range 2 (0

–2 mW/cm

2

), a leakage value of 2 mW/cm

2

will have an analog signal

of 4 VDC and a 1 mW/cm

2

value will have a signal of 2 VDC. On range 4

(0

–10 mW/cm

2

), the 4 VDC analog output signal will indicate 10 mW/cm

2

.

The resistance of the external load connected to the analog output

receptacle should not be less than 5000 ohms.