Peak hold – ETS-Lindgren HI-1710A Microwave Oven Survey Meter User Manual

Page 36

Advertising
background image

36

Menu Options, Settings, Command Set

www.ets-lindgren.com

STIM

—Applies and displays a DC voltage to the input of the RF signal

circuitry; value should be between 2.44 and 4.88. Readings outside

this range indicate a fault in the HI-2623 probe/preamplifier.

TEST

—Voltage at the analog output jack; value should be between

3.80 and 4.20 volts.

If an error occurs in any of the parameter readings, the

parameter screen displays and the erroneous reading

flashes. Turn the HI-1710A off, correct the problem, and

then turn it on again.

PEAK

HOLD

Captures and holds the highest leakage reading displayed.

Menu Option

Setting

Serial Port

Command

ON

P1

OFF

P0

P?

While in Peak Hold the bar graph on the display shows the current

instantaneous RF reading, without filtering. A bar graph segment

remains lit to indicate the relative location of the peak reading. The

bar graph display responds very rapidly and will show stirrer/turntable

modulation as well as minima and maxima of leakage around the

microwave oven periphery.

Because of the fast, unfiltered, response of the bar graph display, the

digital peak reading may lag behind the bar graph indication,

depending on the filter factor chosen. When the instantaneous RF

indication on the bar graph display is greater than the reading on the

digital display, the bar graph segment proportional to the digital reading

is unlit. In this way the Peak Hold mode of the meter can be used to

observe the effects of the various digital filters on fast changing RF

inputs.

In the normal mode (not Peak Hold), the bar graph display follows the

filtered display reading.