Startup Error Messages

As part of the self-test performed on startup, a number of error messages can be

displayed indicating operating problems. In addition, the offset and bias values of

the probe are measured each time the probe is zeroed. If an erroneous reading

is detected, the HI-1710A Microwave Measurement System will display the

parameter screen and the erroneous reading will flash. A message will also be

sent to the RS-232 port. Following is the action to be taken for each error.

REF

C

OUTSIDE

LIMIT

1.24-volt supply low

1.24-volt supply high: The 1.24 volt reference to the A/D converter is

beyond specification. Return for service if error message persists.

+5SUP

C

OUTSIDE

LIMIT

5-volt supply low

5-volt supply high: The 5-volt logic supply is beyond specification.

Voltage is not user-adjustable; return for repair.

+8SUP

C

OUTSIDE

LIMIT

8-volt supply low

8-volt supply high: The 8-volt supply to the analog circuitry is beyond

specification. Voltage is not user-adjustable; return for repair.

-8SUP

C

OUTSIDE

LIMIT

-8-volt supply low

-8-volt supply high: The negative 8-volt supply to the analog circuitry is

beyond specification. Voltage is not user-adjustable; return for repair.