Alarm setpoint, E 30 – ETS-Lindgren HI-1710A Microwave Oven Survey Meter User Manual

Page 30

Advertising
background image

30

Menu Options, Settings, Command Set

www.ets-lindgren.com

ALARM

SETPOINT

Sets a level after which an alarm will sound to indicate that the

measured leakage has exceeded the level.

The alarm will sound only if the audio is turned on. To turn

on the alarm, see Audio on page 31.

Menu Option

Setting

Serial Port

Command

XX.XX

T####

T?

Where #### is a value from 0 to 9999 in mW/cm

2

. For

example, 5000 is 5.0 mW/cm

2

, 5 is 0.005 mW/cm

2

, and so

on.

Each press of the up/down arrow button will raise/lower the setting

by 0.01 for the first ten presses (9.99, 9.98, 9.97, and so on), by 0.10

for the next ten presses (9.89, 9.79, 9.69, and so on), and by 1.00 for

the remaining presses (8.89, 7.89, 6.89, and so on).

Any time a change in direction is made up or down, the change

increment begins at 0.01.

The alarm will sound only when the displayed RF value exceeds the

alarm set point. For example, if the alarm is set at 4.99 mW/cm

2

, the

alarm will only sound when the displayed RF level is 5.00 or greater.

Once the RF level decreases below the alarm set point, the alarm will

turn off.